Laboratory Research Grade Benchtop EC/TDS/Salinity/Resistivity Meter with ATC and Extended Range HI5321-01
Features
- EC Range - 0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC*
- EC Resolution - 0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm
Manufacturer | Hanna Instruments |
---|---|
Hazardous | No |
Note | Ships with heat pack(s) during winter months. |
EC Range | 0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm actual EC* |
EC Resolution | 0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm |
EC Accuracy | ±1% of reading (±0.01 μS/cm) |
EC Calibration | automatic standard recognition (0.000 μS/cm, 84.00 μS/cm, 1.413 mS/cm, 5.000 mS/cm, 12.88 mS/cm, 80.00 mS/cm, 111.8 mS/cm) or user standard; single point or multi-point calibration |
TDS Range | 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt actual TDS* (with 1.00 factor) |
TDS Resolution | 0.001 ppm, 0.01 ppm, 0.1 ppm, 0.001 ppt, 0.01 ppt, 0.1 ppt |
TDS Accuracy | ±1% of reading (±0.01 ppm) |
EC/TDS Temperature Compensation | disabled, linear and non-linear (natural water) |
Temperature Correction Coefficient | 0.00 to 10.00 %/°C |
Resistivity Range | 1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm |
Resistivity Resolution | 0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm* |
Resistivity Accuracy | ±2% of reading (±1 Ω•cm) |
Salinity Range | practical scale: 0.00 to 42.00 psu; natural seawater scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0% |
Salinity Resolution | 0.01 for practical scale/natural seawater scale; 0.1% for percent scale |
Salinity Accuracy | ±1% of reading |
Salinity Calibration | percent scale—one-point (with HI7037 standard) |
Salinity Temperature Compensation | disabled, linear and non-linear (natural water) |
Temperature Range | -20.0 to 120.0˚C**; -4.0 to 248.0˚F**; 253.15 to 393.15 K** |
Temperature Resolution | 0.1 °C, 0.1 °F, 0.1 K |
Temperature Accuracy | ±0.2 °C; ±0.4 °F; ±0.2 K (without probe) |
EC/TDS Probe | HI76312 platinum, four-ring EC/TDS probe with and 1 m (3.3’) cable (included) |
Power Supply | 12 VDC adapter (included) |
Environment | 0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing |
Dimensions | 160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) |
Weight | 1.2 kg (2.64 lbs.) |
Ordering Information | HI5321 is supplied with HI76312 conductivity probe, HI76404W electrode holder, 12 VDC power adapter and instructions |
Special Notes | **Uncompensated conductivity (or TDS) is the conductivity (or TDS) value without temperature compensation. **Reduced to actual probe limits |
Warranty | 2 years |
The HI5321 is an advanced research grade benchtop EC/TDS/Salinity/Resistivity meter that is completely customizable with a large color LCD, capacitive touch keys, and USB port for computer connectivity. The HI5321 is rich in features including 4 point calibration, autoranging, data logging, alarm limits, comprehensive GLP, and many more while retaining simplicity in use with both dedicated key for routine operation and virtual keys that guide the user through setup options.
- Capacitive touch keypad
- Autoranging over a wide range from 0.000 μS/cm to 1000.0 mS/cm
- Preprogrammed with USP method