Copper Half-Cell Ion Selective Electrode (ISE) HI4008
The HI4008 is a solid state, half-cell ion selective electrode (ISE) for the determination of cupric ions (Cu2+) in solution. The mixed cupric sulfide/silver sulfide sensing pellet produces a potential change due to the cupric ion exchange between the inorganic pellet membrane and the sample. The internal sensing elements are housed within a durable epoxy body. The HI4008 is ideal for a variety of applications in plating, wastewater, and industrial processes.
- Solid state sensor
- Detection from 0.065 to 6,355 mg/L Cu2+
- For use with HI5315 reference half-cell
Manufacturer | Hanna Instruments |
---|---|
Electrodes Parameters | Copper |
Electrode Connection | BNC |
Hazardous | No |
Note | Ships with heat pack(s) during winter months. |
Type | solid state; half-cell |
Measurement Range | 0.1M to 1•10-6M 6355 to 0.06 mg/L (ppm) |
Optimum pH Range | 3 to 7 |
Temperature Range | 0 to 80°C |
Diameter | 12 mm |
Length | 120 mm |
Body Material | epoxy |
Cable | 1 m coaxial |
Common Applications | used as an indicator for titrations using chelates |
Connection | BNC |
Warranty | 6 months |
The HI4008 cupric ion selective electrode is a potentiometric half-cell sensor housed within an epoxy body. The internal elements are isolated from the sample by a compressed cupric sulfide/silver sulfide sensing membrane. This solid state sensor requires a separate reference electrode to complete the measurement system.
Hanna’s solid state half-cell ISEs are designed for use with the HI5315 reference half-cell. The reference half-cell features a cone style junction. The skirted cone geometry of the reference forms the liquid junction with the test solution at the tip of the junction cone, producing a highly stable reference potential and consistent flow rates of electrolyte into the solution. A simple press of the top portion of the ISE makes it easy to empty the electrolyte fill solution and replace when necessary.
For the HI4008 to measure cupric ions accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the HI4000-00 ionic strength adjuster (ISA) solution. The ISA ensures that the activity coefficient is constant so the concentration of free cupric ions can be measured. Measuring all solutions with a constant ionic strength reduces the margin of error between measurements. The HI4008 cupric solid state ISE can measure from 1µM (0.065 mg/L) to 0.1M (6,355 mg/L).